2. Place the thermocouple in a known, stable temperature environment (such as an ice bath, lag bath, or a dry well calibrator). 3. Allow the readings to stabilize. 4. In SETUP, change the offset until the primary display reading matches the bath temperature. (See "Changing Setup Options” in the 51/52 Series II Users Manual, and the 53/54 Series II Users Manual.) 51, 52, 53, 54 Series II Service Manual Basic Maintenance Basic maintenance required by the thermometer is discussed in this section. C