Short (or apply 0 mV) to the copper wires at the Calibrator. 5. Verify the thermometer reads 0.0 oC ± 0.3 oC. 6. If this test fails, reposition the thermocouple wires and Mercury thermometer to the same depth in the ice bath, and repeat steps 3-5. If the test continues to fail, perform the calibration procedure, located after this procedure. 7. Remove the short applied in step 4. 8. Using Table 3, steps 2-5, obtain an output from the Calibrator equal to the values in the “Input Voltage (mV DC)”